PRODUCTS
Electrophoresis Power Supplies
* Standard Power Supplies
* Programmable Power Supplies
Electrophoresis Cells
UV Transilluminators
Thermal Cyclers
Ultrasonic Processors
Media Filling / Dosing Pumps
Automated Filling Systems
Pipettes
Block incubators - Dry Baths

Product >> Electrophoresis Power Supplies >> Standard Power Supplies >> ENPS 5000 HV


ELECTROPHORESIS POWER SUPPLIES

Standard Power Supplies


ENPS 5000 HV


Technical Data

ENPS 5000 HV

Output Voltage (V)

20 - 5000 V

Type of Output

Constant Voltage/ Constant Current

Output Current (mA)

0.01-500 mA

Maximum Power (W)

400 W

Number of Output

4

Voltage Setting Resolution

1V UPTO 100 V, 10V from 1000 to 5000 V

Current Setting Resolution

1 m A

Display for Voltage

3Digit

Display for Current

3Digit

Warranty

3 years

Maximum Ripple

1%

Timer

1mnt to 999 mnts

Voltage Regulation (No load to full load)

1%

Maximum Operating Temperature

45°C ambient

Input Supply

230 V AC ± 10%

Weight

3 kg

ENPS 500 PRO - The Universal Programmable power supply

Programming

9 programs each with up to 9 phases

Regulation

Maximum voltage, current and power with automatic crossover at preset limits

Output range

Voltage: 0-500 V Current: 0-2500 mA Power: 0-500 W

Programming range

Voltage: 0-500 V Current: 0-2500 mA Power: 0-500 W Time: 00:01-500 h Volthour: 1-500 000 Vh Amperehour: 1-25 000 mAh

Output resolution

Voltage : 1 V Current : 1 mA

Programming resolution

Voltage : 1V Current : 1mA

Power : 1W

Time: 1min, 00:01-99:59, 1h, 100-500 h

Volthour: 1Vh, 1-9999 Vh100 Vh, 10.0-99.9 kVh

1 kVh, 100-500 kVh Amperehour: 1 mAh, 1-999 mAh

10 mAh, 1.00-9.99 Ah

100 mAh, 10.0-25.0 Ah

Load regulation

< 1% at load change 10-90% of maximum load

Ripple

< 1% at 500 V

Short term stability

< 0.2% /10 h after warm up

Accuracy

Voltage: 1%,± 1 V Current: 1%,±1 mA

Power: 2%, ±1 W

Timer: 0.1%± 1 min, 00:01-99:59 h 0.1%,± 10 min, 100-500 h


Output protection

Fully protected against any overload conditions

Recovery after power failure

Duration< 8 s,

The program continues automatically

Duration 8 s, the program continues after manual restart

Ambient operating temperature

4-40 C

Ambient operating humidity

0-95%

Ambient operating pressure

84106 kPa according to GPR

Mains requirements

220-240 V ~ ; 50/60 Hz

Power consumption

Max 500 W

Line regulation

< 0.2%


Top

 

Copyright 2007 Alliance